NKR-CON-002

Neutral displacement valved connector

Keeps the access closed with no significant fluid movement in the catheter on disconnection. In two versions.

Use for up to 7 days or 200 connections

Neutral displacement valved connector

How it works

ConnectorCatheterVein
On disconnection: No significant fluid movement in the catheter
  1. 1

    Connect

    When a Luer syringe or administration set is attached, the septum opens and fluid flows.

  2. 2

    Disconnect

    On removal, the septum springs back and closes the access. The system stays closed until the next use.

Parts of the connector

Neutral displacement valved connector
  1. 1

    Inlet with septum

    Where the Luer syringe or administration set is attached.

  2. 2

    Valve body

    The split septum opens on connection and closes on disconnection.

  3. 3

    Outlet with protective cap

    Luer connection to the catheter, covered until use.

Product information

Technical characteristics
CharacteristicHigh pressureLow volume
Displacement typeNeutralNeutral
Gravity flow rate (39.3 in head)150 mL/min120 mL/min
Priming volume0.075 mL0.045 mL
Reflux volumeNot stated0.01 mL
Pressure ratingUp to 400 psiUp to 300 psi
Flow rate under pressure2,200 mL/min at 400 psi1,000 mL/min at 200 psi
Leak pressure450 psi320 psi
Functional activations200 over 7 days200 over 7 days
Latex-freeYesYes
MRI compatibleYes, per the manufacturerYes, per the manufacturer
Materials
Upper housing
Tritan MX711 (copolyester) in the high pressure version; polycarbonate in the low volume version. Medical grade.
Valve (septum)
Silicone rubber, medical grade
Lower housing
Polycarbonate, medical grade
Protective cap
Polypropylene
Intended use and limitations
Intended use
Closing the access of the infusion system and allowing needle-free intravenous administration of fluids and drugs, and blood sampling.
Used with
Central venous catheter (CVC), PICC or peripheral venous catheter.
User
Healthcare professionals trained in intravenous therapy.
Limitations
Contraindicated in patients with venous access infarction or severe coagulation disorders. Use in hemodialysis and apheresis is not covered by the manufacturer's instructions. Do not use with needles.
Replacement
Every 7 days or 200 connections, whichever comes first.
Sterilization, shelf life and storage
Sterilization
Ethylene oxide. Sterile and non-pyrogenic.
Shelf life
3 years from sterilization.
Storage
0 °C to 40 °C, 30% to 85% relative humidity, away from direct sunlight.
Use
Single use. Do not reuse or resterilize.
Packaging
Unit
Sterile pouch, 80 × 32 mm
Box
100 units, 210 × 110 × 70 mm

PDF in Portuguese. Instructions for use are sent on request. Contact

Contact Norkor

Quotations, samples and documentation for institutions and distributors.

contato@norkor.com.br